HX
Technical Specifications
|
|
|
HX
75
|
HX
150
|
HX
300
|
|
Standard
Temperature Range
|
|
|
|
|
C
|
5deg
to 35deg
|
5deg
to 35deg
|
5deg
to 35deg
|
|
F
|
45deg to 95deg
|
45deg
to 95deg
|
45deg
to 95deg
|
|
Optional
Temperature Range
|
5deg
to 90deg
|
5deg
to 90deg
|
5deg
to 90deg
|
|
|
(1Kw
heater @ 208VAC)
|
(2.5Kw
heater @ 208VAC)
|
(5Kw
heater @ 208VAC)
|
|
Ambient
Temperature Range
|
|
|
|
|
C
|
13deg
to 35deg
|
13deg
to 35deg
|
13deg
to 35deg
|
|
F
|
55deg
to 95deg
|
55deg
to 95deg
|
55deg
to 95deg
|
|
Stability
|
|
|
|
|
C
|
+/-
0.1deg
|
+/-
0.1deg
|
+/-
0.1deg
|
|
F
|
+/-
0.2deg
|
+/-
0.2deg
|
+/-
0.2deg
|
|
Condenser
|
air
or water cooled
|
air
or water cooled
|
air
or water cooled
|
|
Reservoir
Size
|
5
Gallons/18.9 Liters
|
8
Gallons/30.3 Liters
|
15
Gallons/56.8 Liters
|
|
Cooling
Capacity
|
|
|
|
|
60Hz
at 20degC
|
2,000
W
|
4,500
W
|
10,000
W
|
|
50Hz
at 20degC
|
1,660
W
|
3,735
W
|
8,300
W
|
|
Pump
Performance
|
|
|
|
|
60Hz
Pump 1
|
3.3gpm
@ 50 psig (PD2)
|
3.3gpm
@ 50 psig (PD2)
|
3.3gpm
@ 50 psig (PD2)
|
|
60Hz
Pump 2
|
-2.6gpm
@ 42 psig (TU1)
|
10gpm
@ 40 psig (CP55)
|
9gpm
@ 50 psig (TU5)
|
|
60Hz
Pump 3
|
n/a
|
n/a
|
19gpm
@ 50 psig (CP75)
|
|
50Hz
Pump 1
|
2.75gpm
@ 50 psig (PD2)
|
2.75gpm
@ 50 psig (PD2)
|
2.75gpm
@ 50 psig (PD2)
|
|
50Hz
Pump 2
|
2.5gpm
@ 40 psig (TU1)
|
10gpm
@ 27psig (CP55)
|
8gpm
@ 25psig (TU5)
|
|
50Hz
Pump 3
|
n/a
|
n/a
|
10gpm
@ 40 psig (CP75)
|
|
Power
Requirements
|
|
|
|
|
60Hz
|
208-230V1
diameter/
460V 3
diameter
|
208-230V1diameter/
460V 3
diameter
|
208-230V3diameter/
460V 3
diameter
|
|
50Hz
|
220-240V1diameter/
460V 3
diameter
|
220-240V1diameter/
460V 3
diameter
|
380-415V3diameter/
460V 3
diameter
|
|
Unit
dimensions
|
|
|
|
|
H
x W x D in
|
36.75
x 23.25 x 18.75
|
40.625
x 26.25 x 21.125
|
46.875
x 33.75 x 25.25
|
|
H
x W x D cm
|
93.3
x 59 x 47.6
|
103.2
x 66.7 x 53.7
|
118.9
x 85.7 x 64.1
|
|
Plumbing
Connections
|
|
|
|
|
inlet/outlet
process
|
3/4"
FNPT
|
3/4"
FNPT
|
3/4"
FNPT (CP-75 1" FNPT)
|
|
inlet/outlet
facility (W/C only)
|
1/2"
FNPT
|
1/2"
FNPT
|
1/2"
FNPT
|
|
drain
|
1/2"
FNPT
|
1/2"
FNPT
|
1/2"
FNPT
|
|
auto
refill
|
3/8"OD
SS barb
|
3/8"OD
SS barb
|
3/8"OD
SS barb
|
|
Refrigerant
|
|
|
|
|
60Hz
|
R22
|
R22
|
R22
|
|
50Hz
|
R134A
|
R134A
|
R134A
|
|
Compliance
|
|
|
|
|
50
Hz units
|
CE
|
CE
|
CE
|
|
Unit
Weight
|
|
|
|
|
lb
|
261
|
320
|
477
|
|
kg
|
118.4
|
145.2
|
216
|
|
Standard
Features
|
|
Auto-refill
|
Allows
for self-filling of the chiller to ensure that the proper level
in the reservoir is maintained
|
|
Stainless
steel reservoir
|
Convenient
easy cleaning. Compatible with a wide range of fluids
|
|
Temperature
stability of +/- 0.1¡C
|
Keeps
your process stable giving you consistent, reliable results
|
|
High
and low temperature safeties
|
Can
be configured as warnings or, will shut the unit down to keep your
application safe
|
|
Auto-restart
|
In
the event of power failure, the unit will automatically restart,
upon power restoration which ensures productivity
|
|
Low
level safety
|
Alarms
you if the reservoir level is too low
|
|
Hot
gas by-pass
|
Refrigeration
design that allows for tight temperature stability and longer compressor
life
|
|
Compact
footprint
|
Efficient
design keeping your valuable floor space to a minimum
|
|
15-Pin
analog control port
|
Allows
for remote status of alarms and remote on/off capabilities
|
|
Integrated
fluid pressure gauge and flow control
|
Provides
integral pressure and flow control to adjust to your process needs
|
|
|
|
|
Options
|
|
Pump
selection
|
Various pumps available to meet the flow and pressure requirements
of your application
|
|
Air-cooled
or water-cooled condenser
|
Configurable
to your facilities needs |
|
High
temperature range
|
Allows
heating as well as cooling and high temperature operation up to
+90¡C
|
|
Powerful
TC 400 Controller,
|
User-friendly
interface that allows more sophisticated monitoring and control
of HX operation
|
|
LED status indicators
|
|
|
Alarm Status
|
|
|
Low flow
|
|
|
RS-232
|
|
|
Communication
RS-232
|
Allows
for control of your chiller from your PC or laptop
|
| |
|
|
Accessories
|
|
Remote
temperature probe
|
Allows
for remote temperature control at your application
|
|
Fluid
filtration 5, 25, 40, micron full flow
|
Maintains
particulate-free operating fluid
|
|
Fluid
filtration 5, 25, 40, micron partial flow
|
Maintains
particulate-free operating fluid
|
|
DI
filtration
|
Maintains
a water resistivity level between 1 and 3 megohm/om2 for cooling
applications requiring ultrapure water or electrical isolation of
the application
|
|
Plumbing
package
|
Provides
tubing, insulation and plumbing connections for easy installation
|
|
Condenser
filters
|
Keeps
the condenser clean and your unit performance optimal
|
|
Ethylene
glycol
|
Allows
circulation to temperatures below 8¡C
|