Products - Temperature Control, Process Chillers, Recirculating Chillers, Immersion Coolers, Ultra Low Temperature & Bath Circulators
 


NESLAB HX500

NESLAB HX Series of Process Chillers

The NESLAB HX Series Process Chillers provide you with years of reliable cooling for your critical process applications.

HX Series +5degC to +35degC

Typical applications:
test equipment
military applications
lasers
NMR
s-ray
mass spectrometers
diffusion and cryo pumps
blister packaging
semi-conductors
reactor vessels
MRI
CT
Sputtering
CMP
lineal accelerators
printing
injection moulding


HX Technical Specifications
HX 500
HX 750
Standard Temperature Range
   
C
5deg to 35deg
5deg to 35deg

F

41deg to 95deg
45deg to 95deg
Ambient Temperature Range
C
13deg to 35deg
13deg to 35deg
F
55deg to 95deg
55deg to 95deg
Stability
C
+/- 0.1deg
+/- 0.1deg
F
+/- 0.2deg
+/- 0.2deg
Condenser
air or water cooled
air or water cooled
Reservoir Size
28 Gallons/106 Liters
40 Gallons/ 151.4 Liters
Cooling Capacity
60Hz at 20degC
15,700 W
24,000 W
50Hz at 20degC
13,030 W
19,920 W
Pump Performance

 

60Hz Pump 1
23gpm @ 50 psig (TU9)
23gpm @ 50 psig (TU9)
60Hz Pump 2
19gpm @ 50psig (CP75)
19gpm @ 50 psig (CP75)
50Hz Pump 1
16gpm @ 50 psig (TU9)
16gpm @ 50 psig (TU9)
50Hz Pump 2
10gpm @ 40 psig (CP75)
10gpm @ 40psig (CP75)
Power Requirements
60Hz
208-230V 3diameter/460V 3diameter
208-230V 3diameter/460V 3diameter
50Hz
380-415V 3diameter
380-415V 3diameter
Unit dimensions
   
H x W x D in
51.625 x 46 x 28.75
64.75 x 46 x 29
H x W x D cm
131.1 x 116.8 x 73
164.5 x 116.8 x 73.7
Plumbing Connections

 

inlet/outlet process
1" FNPT
1" FNPT
inlet/outlet facility (W/C only)
1" FNPT
1" FNPT
drain
1/2" FNPT
1/2" FNPT
auto refill
3/8"OD SS barb
3/8"OD SS barb
Refrigerant
   
60Hz
R22
R404A
50Hz
R134A
R404A
Compliance
   
50 Hz units
CE
CE
Unit Weight
   
lb
746
971
kg
338.4
440.4

Standard Features
Auto-refill
Allows for self-filling of the chiller to ensure that the proper level in the reservoir is maintained
Stainless steel reservoir (look at RTE)
Convenient easy cleaning. Compatible with a wide range of fluids
Temperature stability of +/- 0.1degC
Keeps your process stable giving you consistent, reliable results
High and low temperature safeties
Can be configured as warnings or, will shut the unit down to keep your application safe
Auto-restart
In the event of power failure, the unit will automatically restart, upon power restoration which ensures productivity
Low level safety
Alarms you if the reservoir level is too low
Hot gas by-pass
Refrigeration design that allows for tight temperature stability and longer compressor life
Compact footprint
Efficient design keeping your valuable floor space to a minimum
15-Pin analog control port
Allows for remote status of alarms and remote on/off capabilities
Integrated fluid pressure gauge and flow control
Provides integral pressure and flow control to adjust to your process needs
Options
Pump selection
Various pumps available to meet the flow and pressure requirements of your application
Air-cooled or water-cooled condenser
Configurable to your facilities needs
High temperature range
Allows heating as well as cooling and high temperature operation up to +90degC
Powerful TC 400 Controller,
User-friendly interface that allows more sophisticated monitoring and control of HX operation
• LED status indicators
• Alarm Status
• Low flow
• RS-232
Communication RS-232
Allows for control of your chiller from your PC or laptop
   
Accessories
Remote temperature probe
Allows for remote temperature control at your application
Fluid filtration 5, 25, 40, micron full flow
Maintains particulate-free operating fluid
Fluid filtration 5, 25, 40, micron partial flow
Maintains particulate-free operating fluid
DI filtration
Maintains a water resistivity level between 1 and 3 megohm/om2 for cooling applications requiring ultrapure water or electrical isolation of the application
Plumbing package
Provides tubing, insulation and plumbing connections for easy installation
Condenser filters
Keeps the condenser clean and your unit performance optimal
Ethylene glycol
Allows circulation to temperatures below 8degC
 

For a copy of the product brochure or further information, please complete the Information Request Form.

Return to Top of Page I Return to Neslab Range I Return to Product Range I Return to Main Menu

 

 