NESLAB Merlin
Series
|
NESLAB
Merlin Series Recirculating Chillers
The NESLAB Merlin
Series Recirculating Chillers provide a wide range of cooling capacity
all in a highly reliable, space saving unit. They deliver a continuous
supply of cooling capacity up to 5045 watts. Merlin Series 810 -
5045 watts cooling capacity
Typical applications:
lasers
analytical
instrumentation
printing
equipment
electron
microscope
semi-conductor
manufacturing
medial
imaging equipment
reaction
vessels
condensers
|
|
Merlin
Technical Specifications
|
|
Cooling
capacity @ 20°C
|
|
|
|
|
|
|
60
Hz
|
810
watts
|
1250
watts
|
2410
watts
|
3500
watts
|
5045
watts
|
|
|
2765
BTU
|
4266
BTU
|
8225
BTU
|
11946
BTU
|
17219
BTU
|
|
|
697
Kcal
|
1075
Kcal
|
2073
Kcal
|
3010
Kcal
|
4339
Kcal
|
|
50
Hz
|
725
watts
|
1125
watts
|
2180
watts
|
3500
watts
|
4540
watts
|
|
|
2474
BTU
|
3840
BTU
|
7440
BTU
|
11946
BTU
|
15495
BTU
|
|
|
624
Kcal
|
968
Kcal
|
1875
Kcal
|
3010
Kcal
|
3905
Kcal
|
|
Refrigerant
|
R134A
|
R134A
|
R134A
|
R404A
|
R404A
|
|
Temperature
range
|
|
|
|
|
|
|
standard
|
+5°C
to 35°C
|
+5°C
to 35°C
|
+5°C
to 35°C
|
+5°C
to 35°C
|
+5°C
to 35°C
|
|
low
temperature option
|
-15°C
to 35°C
|
-15°C
to 35°C
|
-15°C
to 35°C
|
-15°C
to 35°C
|
-15°C
to 35°C
|
|
Ambient
temperature range
|
|
|
|
|
|
|
C
|
13°
to 35°
|
13°
to 35°
|
13°
to 35°
|
13°
to 35°
|
13°
to 35°
|
|
F
|
55°
to 95°
|
55°
to 95°
|
55°
to 95°
|
55°
to 95°
|
55°
to 95°
|
|
Temperature
stability
|
+/-
0.1°C
|
+/-
0.1°C
|
+/-
0.15°C
|
+/-
0.15°C
|
+/-
0.15°C
|
|
Pumps
|
|
|
|
|
|
|
MD
30 - Mag Drive*
|
yes
|
yes
|
n/a
|
n/a
|
n/a
|
|
PD
1 - Positive Displacement
|
yes
|
yes
|
yes
|
n/a
|
n/a
|
|
PD
2 - Positive Displacement
|
n/a
|
yes
|
yes
|
yes
|
yes
|
|
CP
55 - Centrifugal Pump
|
n/a
|
n/a
|
n/a
|
yes**
|
yes**
|
|
Reservoir
volume
|
|
|
|
|
|
|
gallon
|
0.5
|
0.5
|
0.5
|
0.5
|
0.5
|
|
liter
|
1.8
|
1.8
|
1.8
|
1.8
|
1.8
|
|
Unit
dimensions
|
|
|
|
|
|
|
H
x W x D in
|
23.50
x 12.62 x 20.85
|
23.50
x 12.62 x 20.875
|
26.125
x 16.25 x 24
|
30.375
x 21.25 x 29.25
|
30.375
x 21.25 x 29.25
|
|
H
x W x D cm
|
59.7
x 32.1 x 53.0
|
59.7
x 32.1 x 53.0
|
66.4
x 41.3 x 61.0
|
77.2
x 54.0 x 74.3
|
77.2
x 54.0 x 74.3
|
|
Power
requirements
|
|
|
|
|
|
|
option
1
|
115V,
60 Hz,
10 amps
|
115V,
60 Hz,
13 amps
|
200-230V,
60 Hz,
9 amps
|
208-230V,
60 Hz,
13 amps
|
208-230V,
60 Hz,
15 amps
|
|
|
100V,
50 Hz,
10 amps
|
100V,
50 Hz,
13 amps
|
200V,
50 Hz,
9 amps
|
200V,
50 Hz,
13 amps
|
200V,
50 Hz,
15 amps
|
|
option
2
|
230V,
50 Hz,
5 amps
|
200-230V,
60 Hz,
7 amps
|
230,
50 Hz,
9 amps
|
230V,
50 Hz,
13 amps
|
230V,
50 Hz,
16 amps
|
|
|
|
200V,
50 Hz,
7 amps
|
|
|
|
|
option
3
|
n/a
|
230,
50 Hz,
7 amps
|
n/a
|
400V,
50 Hz, 3 ph
8 amps
|
400V,
50 Hz, 3 ph
8 amps
|
|
Unit
weight
|
|
|
|
|
|
|
lb
|
118
|
120
|
176
|
254
|
254
|
|
kg
|
54
|
54
|
80
|
115
|
115
|
|
Compliance
|
|
|
|
|
|
|
60
Hz
|
UL/cUL,
CE
|
UL/cUL,
CE
|
UL/cUL,
CE
|
UL/cUL,
CE
|
UL/cUL,
CE
|
|
50
Hz
|
CE
|
CE
|
UL/cUL,
CE
|
CE
|
CE
|
|
Standard
Features
|
|
Variety
of pumps
|
Meets
your specific application needs
|
|
Refrigeration
system
|
Provides
CFC-free refrigeration system for precise temperature control and
optimum stability
|
|
Hot
gas bypass
|
Extends
chiller life
|
|
Reservoir
isolation valves
|
Eliminates
reservoir overflow when pumping vertically
|
|
Fluid
reservoir
|
Enables
you to view the liquid level and fluid condition at a glance
|
|
Removable
front grille
|
Allows
quick release of front grille for routine cleaning
|
|
Integrated
fluid pressure and flow adjustment
|
Meets
specific application needs
|
|
Easy
access reservoir drain
|
Allows
fast and easy access when changing fluids
|
|
Pressure
gauge
|
Provides
fluid system diagnostics at a glance
|
|
Digital
temperature controller
|
Provides
intuitive controller with an easy-to-view digital display for simple
operation
|
|
High/low
temperature safeties
|
Ensures
your application maintains critical operating parameters
|
|
Compact
design
|
Maximizes
valuable floor space
|
|
Two-year
warranty
|
Provides
you with peace of mind
|
| |
|
|
Options
|
|
Communications
package:
|
|
|
RS-232 or RS-485
|
Allows
for remote operation, monitoring, and data logging
|
|
Remote start/stop
|
Allows
the unit to be remotely turned on and off for convenient operation
|
|
Status relay
|
Allows
for remote monitoring
|
|
Remote sensor port
|
Allows
remote temperature control of the application
|
|
Low
level/low flow safety package
|
Ensures
fluid level and fluid flow safety to ensure continuous operation
|
|
|
|
|
Accessories
|
|
NESLAB
NEScom software
|
Allows
you to easily program and automate your temperature control process
from a PC system |
|
Plumbing
kits
|
|
|
NESLAB
NEScom software
|
Allows
circulation between -25°C and +100°C and includes 25' of
tygon tubing, 25' of tubing insulation, and 4 hose clamps
|
|
Plumbing
kits
|
Provides
tubing, insulation, and connectors for easy installation
|
|
External
temperature gauge
|
Allows
you to monitor system operating temperature any point in-line between
the chiller and the application
|
|
External
pressure gauge
|
Allows
you to monitor system pressure any point in-line between the chiller
and the application
|
|
Air
filter kits
|
Protects
your chiller from dusty environments
|
|
Deionized
water package
|
Maintains
a water resistivity level between 1 and 3 megohm/cm2 for cooling
application requiring ultra pure water
|
|
Fluid
particulate filters
|
Maintains
particulate-free operating fluid. Available with 5, 10, 25 or 40
micron filters
|
|
Ethylene
glycol
|
Allows
circulation to temperatures down to -20°C in a 50/50 mixture
with water
|
|