The
ViscoScope VA-300S
The ViscoScope VA-300S sensor a custom designed sensor, which is
specifically constructed to measurement requirements a particular
customer may have. This sensor is available for temperatures up
to 1500degC.
Range:
1.0 to 2,500,000 mPa.s x gr/cm3
Typically
the VA-300S sensor is calibrated for a three decade range but is
capable to cover more than 4 decades of viscosity measurement.
The
ViscoScope VA-300S sensor generates a very small torsional oscillation
resonance of constant amplitude on the probe surface. By adjusting
the amplitude, the sensor can even be configured for difficult applications.
A closed PID loop controls the amplitude of system. Viscosity is
being measured by extracting the viscosity signal from the PID control
loop. Using well established as well as new PID control technology
enables a large measurement range, long term stability and fast
response.
|
The entirely
welded construction seals the sensor from the process. By utilizing an
extremely small amplitude of torsional resonance motion and a sealed construction
the internals of the sensor are not exposed to any wear and tear.
The ViscoScope
VA-300S sensor can be mounted in any position into a tank, reactor or
pipe. For certain applications, which require it, it is also possible
to mount the sensor in an open channel. The measurement can be made, where
it is most important for the application. Current process parameters are
made available for measurement and control purposes. Turbulent flow in
a tank, reactor or pipe does not have an influence on the measurement.
Maintenance
free operation of the ViscoScope sensor VA-300S provides low cost of ownership.
The sensor
is available for three different temperature ranges:
VA-300S-LT: Maximum Process temperature 130degC / 270degF
VA-300S-MT: Maximum Process temperature 300degC / 570degF
VA-300S-HT: Maximum Process temperature 450degC / 840degF
VA-300S-XT: Maximum Process temperature 1500degC / 2700degF
Dependent
on the temperature classification, the temperature is measured with one
or two Pt100 temperature sensors. The process temperature measurement
is being done directly in the viscosity probe and at the same location
at which the viscosity is being measured. A second Pt100 monitors the
coil temperature.
Use of
the sensor in a hazardous, classified area is accomplished via safety
barriers. The sensor can be provided with different agency approvals.
The ViscoScope
VA-300S sensor is compatible with the ViscoScope electronics VS-4450 and
VS-B450.
|
Technical Specifications
|
|
Sensor
Dimensions
|
Dependent
on construction
|
|
Viscosity
range in mPa.s x g/cm3
|
0.10
to 2,500,000
|
|
Calibration
|
3 Decades (optional 4 Decades)
Computer assisted calibration with NIST traceable oils
|
|
Wetted
parts 1) / housing
|
316
(1.4571) / IP65
|
|
Process
temperature 2)
|
LT
-40 to + 130 dec C / -40 to +270 deg F
ST -40 to + 300 dec C / -40 to +570 deg F
HT -40 to + 450 dec C / -40 to +840 deg F
XT -40 to + 1,500 dec C / -40 to +2,750 deg F
|
|
Process
connection 3)
|
Flange:
ASME 3" 300# or DN80 PN40 std.
|
|
Process
pressure
|
Dependent
on flange, max. 350 bar / 5,000 psig
|
|
Resonance
frequency
Shear rate
|
Dependent
on construction
Dependent
on construction
|
|
Transmission
cable length
Sensor to transmitter
|
Maximum
1,000 meters / 3,333 feet
(short cable length recommended for very low viscosities)
|
|
Speed
of flow
|
Up
to 10m / sec or 33 feet / sec dependent on installation
|
|
Reproductivity
|
+/-
1% of reading or +/- 1 Digit
OPTIONAL +/- 0,5% of reading or +/- 1 Digit
|
|
Accuracy
|
+/-
2% of reading or +/- 1 Digit
OPTIONAL +/- 1% of reading or +/- 1 Digit
|
|
OPTIONS
|
|
Non-active-extension
4)
|
Eliminates
no flow areas in a pipe connection on a reactor or in a T-piece.
Can also be used to bridge gaps in open channel applications.
Diameter = 45mm / 1.75", L = up to 100mm / 4.0"
|
|
Hazardous
area approved
|
ATEX:
II 1/2 G EEx ia IIC T3 - T6
Others on request
|
1) Hastelloy C22,
Duplex 2205, Teflon coating and other materials on request
2) Above 175 deg C / 350 deg F coil temperature air colling of the housing
is required
3) Flanges up to PN250 und 2500# ASME on request
4) Non-active extensions are available in various diameters and up to
500mm / 20" length
(*) Subject to change
without notice.
For
a copy of the product brochure or further information, please complete
the Information Request Form.
Return
to Top of Page I Return to Process Viscometers I Return
to Product Range I Return
to Main Menu
|