Products - Sensor Systems for HAAKE Viscometers & Rheometers
 

Sensor Systems for HAAKE Viscotester 550

The modular design of the HAAKE Viscotester 550 system allows the use of nearly any known sensor. For example
Coaxial Cylinder sensors according to DIN 53018 and ISO 3219. Exact temperature control can be provided by use of a circulator
Immersion sensors according to DIN 53019 and ISO 3219
Cone-and-plate sensors according to ISO 3219 and parallel plate sensors
Relative sensors according to ISO 2555
Special immersion sensors for highly filled samples or containing large particles, for the determination of the yield point

All these sensors guarantee that your results will be absolutely comparable either within your company or even worldwide with other laboratories.

 

 

Below is a selection of commonly used sensor systems. For other sensor systems please contact us.

Standard Technical Specifications
Sensor
NV
MV1
MV2
MV DIN
SV1
SV2
SV DIN
Cup
NV
MV
MV
MV/TMV
SV
SV
SV/TSV
Temperature Vessel
+
+
+
+/-
+
+
+/-
Application
Low
viscosity
Medium viscosity
liquids
High viscosity
liquids
Vscosity Range (mPas)
2-2000
3-10000
8-30000
8-18000
50-10(5)
100-3 . 10(5)
50-10(5)
Recom. Viscosity Range (mPas)
5-300
10-7000
20-20000
15-10000
150-60000
300-2 . 10(5)
150-60000
Shear Rate (s-1)
27-27000
11.7-1170
4.5-450
6.5-645
4.5-445
4.5-445
6.5-645
Sample Volume (cm3)
9
40
55
50
9
6
14
Gap (mm)
0.35
0.96
2.6
1.64
1.45
1.45
0.9
Radius, Length (mm)
20.1/60
20.04/60
18.4/60
19.36/58.08
10.1/61.4
10.1/19.6
10.65/31.95
Repeatability (± %)
1
1
1
1
1
1
1
Reproducibility (± %)
3
2
2
3
3
3
3

 

 

Below is a selection of commonly used sensor systems. For other sensor systems please contact us.

Special Technical Specifications
Sensor
PK 1,1¡
PK 1,0.5¡
PK 2,1¡
PK 2,05¡
HV 2 DIN
MV 2P
SV 2P
Cup
-
-
-
-
HV 2
MVP
SVP
Temperature Vessel
PK 100 D or PK 200
+
+
+
Application
Small sample volume, high viscosity samples
High viscosity
Slippage
Vscosity Range (mPas)
50-10(5)
20-50000
100-3 .10(5)
100-1.5.10(5)
103-2-10(6)
8-30000
100-3.5-10(5)
Recom. Viscosity Range (mPas)
100-70000
50-40000
200-2.10(5)
150-800000
2-103-10(6)
20-20000
350-2.10(5)
Shear Rate (s-1)
30-3000
60-6000
30-3000
60-6000
6.5-645
4.4-440
4-390
Sample Volume (cm3)
0.1
0.1
<0.1
<0.1
1
55
6
Gap (mm)
0.0174 rad
0.0087 rad
0.0174 rad
0.0087 rad
0.34
2.6
1.45
Radius, Length (mm)
14/-
14/-
10/-
10/-
4/11.8
18.4/60
10.1/19.6
Repeatability (± %)
2
2
3
3
2
2
2
Reproducibility (± %)
4
4
6
6
4
4
5

 

 

Below is a selection of commonly used sensor systems. For other sensor systems please contact us.

Immersion Technical Specifications
Sensor
E3
E30
E100
E500
E1000
FL10
FL100
FL1000
Cup
Cup Beaker and sample volume depend on application
Adaptor
Adapter is required (part.-no 808-0579 or 222-1359)
Application
Fast and simple relative measurements
Relative measurement
of disperse samples
Temperature
-30 - 200degC
-30 - 200degC
Viscosity (mPas)
3-10(2)
10(2)-10(5)
10(3)-10(5)
5.10(3)-5.10(6)
10(4)-10(7)
10(2)-10(5)
10(3)-10(6)
10(4)-10(7)
Radius, Length (mm)
25/116
12/50.5
8/34.5
5/9
3.5/17.7
20/60
11/16
5/8.8
Repeatability (± %)
3
3
3
5
5
3
5
5

 

 

Below is a selection of commonly used sensor systems. For other sensor systems please contact us.

ISO/ASTM Technical Specifications
Sensor
B1
B2
B3
B4
B5
B6
B7
Cup
Beaker and sample volume depend on application
Adaptor
ISO Adapter is required part no. 222-1204
Application
Viscosity measurement according to ISO 2555, ASTM D 115-72, D789-73, D2196-68
Temeprature
-30 - 200degC
Viscosity (mPas)
10(2)-10(5)
500-5.10(5)
10(3)-10(6)
2.10(3)-2.10(6)
3.10(3)-5.10(6)
8.10(3)-10(7)
3.10(4)-5.10(7)
Radius, Length (mm)
28.13/22.5
23.5/1.65
17.35/1.65
13.65/1.65
10.55/1.65
7.3/1.65
1.6/50.4
Repeatability (± %)
3
4
5
5
5
5
7


 

 

For a copy of the product brochure or further information, please complete the Information Request Form.

Return to Top of Page I Return to Sensor Systems I Return to Viscotester Range I Return to Product Range I Return to Main Menu

 

 

Tel: 03 5367 7477 Fax: 03 5367 6477 Email: info@rheologysolutions.com